The electrolytic cell is upgraded on the basis of C008-10, adding a light window
• The light window of the electrolytic cell adopts quartz with light transmittance greater than 95%
• The light window part of the electrolytic cell is linked with the cell body by a lock ring to ensure sealing and facilitate the replacement of the quartz light window.
• The cell is absolutely sealed system can be vacuumed through the valve (less than -100Kpa), the cell can be slightly pressurized in order to
For safety reasons, the pressure should not be greater than 0.5MPa. • Due to process restrictions, the inner wall glass of the double-layer electrolytic cell is relatively weak, so the pressure is prohibited
• The electrodes of the cell must be customized to match the cell.
• The reference electrode of the electrolytic cell is separated from the working electrode and connected by the Lukin capillary tube to effectively reduce R drop.
• The pore size of the tip part of the Lukin capillary extending into the working electrode chamber of the electrolytic cell shall not be greater than 0.2mm.
• AGAR can be added to the Lukin capillary of the electrolytic cell to reduce the flow rate of the electrolyte.
• The reference electrode chamber of the electrolytic pool is sealed with internal thread, which must be sealed when used to effectively avoid the reference electrode chamber and the working electrode chamber
Leakage due to the communicator effect.
• The electrolytic cell is often used in three-electrode system, the volume of the electrolytic cell is from 30ml to 500ml, and can also be customized according to requirements.
The electrolytic cell can add a sampling port on the cell body, in which case additional costs are required.
No valve can be added to the electrolytic cell.
All absolute seals are realized by quick-connect.
The diameter of the PTFE tube used for quick-connect is 3mm.